Adult Leg for X-Ray CT, US, MRI
Adult Leg for X-Ray/CT, US, and MRI has a joint at the knee for flexibility and motion. It connects the lower leg with the upper leg (thigh). These two parts can be disarticulated as per needs. Our patented bone mimicking material allows for realistic X-Ray CT, US, and MRI images.
This X-Ray/CT, US and MRI Adult Leg Phantom can be used to take X-Ray images of the full leg. The model includes synthetic bones made out of realistic patented bone-mimicking material. The X-Ray Leg Model is perfect for radiology schools for training and learning different settings of X-Ray without the danger of exposing an actual patient to harmful radiation. The leg phantom has a joint at the knee for flexibility and motion. It connects the lower leg with the upper leg (thigh).
Design




X-Ray Scans





CT Scans


MRI Scans


- Thigh (Femur)
- Knee Joint
- Leg (Tibia, Fibula)
- Foot (Toe Bones)
| Phantom Size | 38 x 10 x 7 Inches |
| Phantom Weight | 27 Lbs (approx.) |
| Shipment Size | 36 x 24 x 12 Inches |
| Shipment Weight | 35 Lbs (approx.) |
| Type of Tissue | Sound Velocity [m/s] | Density [g/cm³] | Hardness [Shore 00] | T2 [ms] | Hounsfield Value | Attenuation Measured at 2.25 MHz [dB/cm] |
| Cortical Bone | 3000 ± 30 | 2.31 | N/A | N/A | 1300 | 6.4 ± 0.3 |
| Trabecular Bone | 2800 ± 50 | 2.03 | N/A | N/A | 1300 | 21 ± 2 |
| Body Tissue | 1410 ± 10 | 1.02 | 60 | 50 ms | 25 | 1.7 ± 0.2 |
| Thermal Properties of Bone-Mimicking Material | |||||
| Thermal Conductivity | Volumetric Specific Heat Capacity | Thermal Diffusivity | Thermal Resistivity | Specific Heat | Speed of Sound |
| 0.776 W/ m K | 1.040 MJ/ m^3 K | 0.746 mm^2/ s | 1.289 m K/ W | 0.978 J/ g Deg Celsius | 3070 m/s |
| HU Values of the Tissue-Mimicking Materials | ||
| S.No. | Tissue Type | HU Value (average) |
| 1 | Trabecular Bone | 800 |
| 2 | Cortical Bone | 1300 |
| Leg Model | Leg (Default – Bones Only), Leg (With Muscles) |