X-Ray, CT, US, MRI Phantoms
Newborn Skull for X-Ray CT, US, MRI
$2,900 USD
The Newborn Skull Phantom serves for medical imaging research and treatment planning of non-invasive HIFU brain surgeries. It replicates an average newborn skull using a realistic patented epoxy-based bone material suitable for Ultrasound, MRI, and X-Ray/CT.
This product can be customized based on your requirements — contact us.
The Newborn Skull Phantom serves for medical imaging research and treatment planning of non-invasive HIFU brain surgeries. It replicates an average newborn skull using a realistic patented epoxy-based bone material suitable for Ultrasound, MRI, and X-Ray/CT.
Design




CT Scans

- Cadaveric Specimen Phantom Specimen
- Based on an Average Newborn Human Skull
- Realistic Bone-Mimicking Material
- A Wydra, RG Maev, Phys Med Biol, Vol 58(22),
| Phantom Size | 4 x 6 x 5 Inches |
| Phantom Weight | 2 Lbs (approx.) |
| Shipment Size | 12 x 10 x 10 Inches |
| Shipment Weight | 6 Lbs (approx.) |
| Type of Tissue | Sound Velocity [m/s] | Density [g/cm³] | Hardness [Shore 00] | T2 [ms] | Speckles | Attenuation Measured at 2.25 MHz [dB/cm] |
| Cortical Bone | 3000 ± 30 | 2.31 | N/A | N/A | N/A | 6.4 ± 0.3 |
| Thermal Properties of Bone-Mimicking Tissue | |||||
| Thermal Conductivity | Volumetric Specific Heat Capacity | Thermal Diffusivity | Thermal Resistivity | Specific Heat | Speed of Sound |
| 0.776 W/ m K | 1.040 MJ/ m^3 K | 0.746 mm^2/ s | 1.289 m K/ W | 0.978 J/ g Deg Celsius | 3070 m/s |
| HU Values of the Tissue-Mimicking Materials | ||
| S.No. | Tissue Type | HU Value (average) |
| 1 | Cortical Bone | 1300 |